Ophir Temperature Ranges for Calibration

Modified on Tue, 6 Jan at 8:14 AM

Acceptable ambient temperature ranges for Ophir products vary by product type:


Displays - 19-28C

Thermal Sensors - 17-32C

Pryo-Electric Sensors - 17-32C

Photodiode Sensors -  19-31C

Wavelength Sensors - 19-31C


The ambient temperature is controlled in each calibration lab and is monitored by our calibration software. If the ambient temperature in a lab is ever outside of the acceptable range the calibration software will not allow for calibration.